This video will feature the benefits and proper use of UltraSnap, Hygiena’s best-selling ATP sanitation monitoring test.
2
UltraSnap is a user-friendly ATP test used for assessing the cleanliness of food contact surfaces and other processing equipment after cleaning.
3
Together, UltraSnap and Hygiena’s hand held meters measure the ATP present in a sample.
4
ATP is the energy molecule found in all living cells and the presence of ATP signifies whether or not a surface is clean.
5
Test results are delivered in 15 seconds and are expressed in Relative Light Units or RLUs.
6
An RLU result is directly proportional to the amount of ATP collected from the sample, so the greater the amount of ATP, the greater the RLU number, and the dirtier the surface.
7
UltraSnap’s pre-moistened swab bud breaks through biofilms and ensures maximum sample collection for accurate results.
8
The liquid-stable chemistry in Hygiena’s patented Snap-Valve is facilitates superior sensitivity and reliable results.
9
UltraSnap has a shelf life of 15 months at refrigerated temperatures and 4 weeks at room temperature.
10
Turn on the luminometer before you begin testing to allow time for the systems calibration self-check to run.
11
If you've already programmed test locations into your luminometer with SureTrend software, select the location to be tested.
12
Remove the swab from the tube.
13
Swab a 4x4 in (10x10cm) square area in a criss-cross pattern, swabbing the area with even coverage.
14
Apply sufficient pressure to maximize sample collection and improve the accuracy of test results.
15
Be sure not to touch the swab shaft when collecting sample.
16
This could contaminate the test and cause inaccurate results.
17
Rotate the swab tip while swabbing to ensure maximum sample collection.
18
When swabbing an irregular surface, ensure the swabbing technique remains consistent and swab a large enough area to collect a representative sample.
19
Replace the swab in the tube.
20
The easiest way to activate UltraSnap is to hold the device in your fist and use the thumb and forefinger to break the Snap-Valve by bending the bulb forward and backward.
21
Next, squeeze the bulb twice to expel liquid into the tube.
22
Then shake for 5 seconds.
23
Insert the swab into the read chamber and close the lid.
24
Hold the luminometer upright and press OK to initiate measurement.
25
Results will be displayed in 15 seconds.
26
Find more information about setting appropriate RLU limits for your facility at www.hygiena.com or by watching the Setting Pass/Fail RLU video on our youtube channel.
27
Since UltraSnap is made of 100% recyclable plastic, used devices may be discarded in the recycle bin.
28
To get the most out of your monitoring system, use Hygiena’s complimentary SureTrend software to track and trend your test results.
29
SureTrend software provides invaluable access to preset and customizable reports, making it easy to recognize trends and analyze sanitation effectiveness.
30
SureTrend also assists in identifying environmental contamination, verifies sanitation as part of a HACCP program, and maintains records for audit compliance.
31
For more information about UltraSnap or any other Hygiena product, contact us with the information on the screen.
32
With technical staff and distributors around the world, all Hygiena products come with personal account managers, online resources, and 24/7
下载应用
Everything you need to speak fluently
AI PronunciationScore every sentence
IPA PracticeMaster every sound
VocabularyBuild your word bank
Vocab GameLearn while playing
关于本课
您正在使用跟读技巧通过视频"UltraSnap® Surface ATP Test"练习英语口语和发音。
每天练习15到30分钟,将显著提高您的英语流利度和发音准确度。
什么是跟读法?
跟读法 (Shadowing) 是一种有科学依据的语言学习技巧,最初开发用于专业口译员的培训,并由多语言者Alexander Arguelles博士普及。这个方法简单而强大:您在听英语母语原声的同时立即大声重复——就像是一个延迟1-2秒紧跟说话者的影子。与被动听力或语法练习不同,跟读法强迫您的大脑和口腔肌肉同时处理并模仿真实的讲话模式。研究表明它能显着提高发音准确性,语调,节奏,连读,听力理解和口语流利度——使其成为雅思口语备考和真实英语交流最有效的方法之一。